High Range NTU Digital Turbidity Sensor

Short Description:

★ Model No: ZDYG-2088-01QX

★ Protocol: Modbus RTU RS485 

★ Power Supply: DC12V

★ Features: Scattered light principle, automatic cleaning system

★ Application: Sewage water, ground water, river water, water station


    • facebook
    • sns02
    • sns04

    Product Detail

    Manual

    Brief Introduction

    The ZDYG-2088-01QX Turbidity sensor is based on the infrared absorption scattered light method and combined with the application of ISO7027 method, can guarantee the continuous and accurate detection of suspended solids and sludge concentration. Based on ISO7027, infrared double scattering light technology will not be affected by chroma for the measurement of suspended solids and sludge concentration value. According to the usage environment, self-cleaning function can be equipped with. It assures the stability of data and reliability of performance; with the built-in self-diagnosis function.The digital suspended solid sensor measure the water quality and delivery the data in high precision,the sensor installation and calibration is quite simple too.

    Application

    Widely using in sewage plant, water plant, water station, surface water, farming, industry and other fields.

    ZDYG-2087-01QX      /zdyg-2087-01qx-online-sludge-concentration-sensor-product/          污染水源1

    Technical Parameters

    Measurement Range

    0.01-100 NTU ,0.01-4000 NTU

    Communication

    RS485 Modbus

    Main Materials

    Main Body:SUS316L (Ordinary Version),Titanium Alloy (Seawater Version)

    Upper and Lower Cover:PVC

    Cable: PVC

    Waterproof Rate

    IP68/NEMA6P

    Indication Resolution

    Less than ± 5% of the measured value (depending on sludge homogeneity)

    Pressure Range

    ≤0.4Mpa

    Flow velocity

    ≤2.5m/s,8.2ft/s

    Temperature

    Storage Temperature:-15~65℃;Environment Temperature:0~45℃

    Calibration

    Sample Calibration, Slope Calibration

    Cable Length

    Standard 10-Meter Cable, Max Length: 100 Meters

    Power Supply

    12 VDC

    Warranty

    1 year

    Size

    Diameter 60mm* Length 256mm

     

    Wire connection of Sensor

    Serial No. 1 2 3 4
    Sensor Cable Brown Black Blue White
    Signal +12VDC AGND RS485 A RS485 B

     