We now have our individual sales group, layout team, technical team, QC crew and package group. Now we have strict high-quality control procedures for each procedure. Also, all of our workers are experienced in printing discipline for Tds Ph And Conductivity Meter , Ph Probe , Ph Electrode , Our firm warmly welcome friends from everywhere in the globe to visit, examine and negotiate business enterprise.
Wholesale China Dissolved Oxygen Analyzer Factory Suppliers - DOG-208F Industrial Dissolved Oxygen Sensor – BOQU Detail:

DOG-208F Dissolved Oxygen Electrode applicable for Polarography Principle

With platinum (Pt) as cathode and Ag / AgCl as anode
The electrolyte is 0.1 M potassium chloride (KCI).
The silicone rubber permeable membrane imported from U.S. serves as the permeable
membrane.
It has silicone rubber and steel gauze.
It is featured by collision resistance, corrosion resistance, high.temperature resistance, shape
retention and other performances
Measuring range: 0-100ug/L 0-20mg/L
Electrode material: 316L stainless steel
Temperature compensation resistor: 2.252K 22K Ptl00 Ptl000 etc
Sensor life: >3 years
Cable length: 5m (double shielded)
Detection lower Limit: 0.1ug/L(ppb)(20℃)
Measurement upper limit: 20mg/l(ppm)
Response time: ≤3min(90%, 20℃)
Polarization time: >8h
Minimum flow rate: 5cm/s; 515 L/h
Drift: <3%/month
Measurement error: <±1 ppb
Air current: 50-80nA   Note: Maximum current 20-25 uA
Polarization voltage: 0.7V
Zero Oxygen: <5ppb(60min)
Calibration intervals: >60 days
Measured water temperature:0~60℃
Measuring range: 0-100ug/L 0-20mg/L
Electrode material: 316L stainless steel
Temperature compensation resistor: 2.252K 22K Ptl00 Ptl000 etc
Sensor life: >3 years
Cable length: 5m (double shielded)
Detection lower Limit: 0.1ug/L(ppb)(20℃)
Measurement upper limit: 20mg/l(ppm)
Response time: ≤3min(90%, 20℃)
Polarization time: >8h
Minimum flow rate: 5cm/s; 515 L/h
Drift: <3%/month
Measurement error: <±1 ppb
Air current: 50-80nA   Note: Maximum current 20-25 uA
Polarization voltage: 0.7V
Zero Oxygen: <5ppb(60min)
Calibration intervals: >60 days
Measured water temperature:0~60℃

 

Applied to thermal power plants, power plant desalted water,.

boiler feed water etc places of trace oxygen content.


Product detail pictures:

Wholesale China Dissolved Oxygen Analyzer Factory Suppliers - DOG-208F Industrial Dissolved Oxygen Sensor  – BOQU detail pictures


Related Product Guide:

Persisting in "High quality, Prompt Delivery, Competitive Price", we have established long-term cooperation with clients from both overseas and domestically and get new and old clients' high comments for Wholesale China Dissolved Oxygen Analyzer Factory Suppliers - DOG-208F Industrial Dissolved Oxygen Sensor – BOQU , The product will supply to all over the world, such as: Uzbekistan, Kazakhstan, Seychelles, With the technology as the core, develop and produce high-quality products according to the diverse needs of the market. With this concept,the company will continue to develop products with high added values and continuously improve products,and will provide many customers with the best products and services!
  • The sales manager is very patient, we communicated about three days before we decided to cooperate, finally, we are very satisfied with this cooperation!
    5 Stars By Marina from Mexico - 2018.11.06 10:04
    Goods just received, we are very satisfied, a very good supplier, hope to make persistent efforts to do better.
    5 Stars By Ann from Kuala Lumpur - 2017.04.28 15:45
    Write your message here and send it to us