DPD Colorimetry Chlorine Analyzer CLG-6059DPD

Short Description:

DPD Colorimetry Chlorine Analyzer CLG-6059DPD
This product is a DPD residual chlorine online analyzer independently developed and manufactured by our
company. This instrument can communicate with PLC and other devices through RS485 (Modbus RTU
protocol), and has the characteristics of fast communication and accurate data.
Application
This analyzer can automatically detect the residual chlorine concentration in water online. The reliable
national standard DPD colorimetric method is adopted, and the reagent is automatically added for
colorimetric measurement, which is suitable for the monitoring of residual chlorine concentration in the
process of chlorination and disinfection and in the drinking water pipe network.
Features:
1) Wide power input, touch screen design.
2) DPD colorimetric method, the measurement is more accurate and stable.
3) Automatic measurement and automatic calibration.
4) The analysis period is 180 seconds.
5) The measurement period can be selected: 120s~86400s.
6) You can choose between automatic or manual mode.
7) 4-20mA and RS485 output.
8) Data storage function, support U disk export, can view historical and calibration data.
Product Name Online Chlorine Analyzer
Measurement principle DPD colorimetry
Model CLG-6059DPD
Measurement Range 0-5.00mg/L(ppm)
Precision Choose the larger of ±5% measurement value or ±0.03 mg/L(ppm)
Resolution 0.01mg/L(ppm)
Power Supply 100-240VAC, 50/60Hz
Analog Output 4-20mA output,Max.500Ω
Communications RS485 Modbus RTU
Alarm Output 2 relay ON/OFF contacts, independent setting of Hi/Lo alarm points,with hysteresis setting,5A/250VAC or 5A/30VDC
Data Storage Data storage function, support U disk export
Display 4.3inch color LCD touch screen display
Dimensions/Weight 500mm*400mm*200mm(Length * width * height); 6.5KG (No reagents)
Reagent 1000mLx2, about 1.1kg in total; can be used about 5000 times
Measure Interval 120s~86400s; default 600s
Single measurement time About 180s
Language Chinese/English
Operating Conditions Temperature: 5-40℃
Humidity: ≤95%RH (non-condensing)
Pollution: 2
Altitude: ≤2000m
Overvoltage: II
Flow rate: 1L/min is recommended
Operating conditions Sample flow rate:250-300mL/min,Sample inlet pressure:1bar(≤1.2bar)
Sample temperature:5~40℃


  • DPD Colorimetry Chlorine Analyzer: