Our items are commonly identified and trusted by customers and may fulfill continuously switching economic and social wants of Ph Sensor , Tds Ppm Meter , Water Treatment Ph Sensor , We welcome new and previous clients from all walks of lifestyle to get in touch with us for long term business relationships and mutual accomplishment!
China Wholesale Online Dissolved Oxygen Meter Manufacturers Pricelist - DOG-208F Industrial Dissolved Oxygen Sensor – BOQU Detail:

DOG-208F Dissolved Oxygen Electrode applicable for Polarography Principle

With platinum (Pt) as cathode and Ag / AgCl as anode
The electrolyte is 0.1 M potassium chloride (KCI).
The silicone rubber permeable membrane imported from U.S. serves as the permeable
membrane.
It has silicone rubber and steel gauze.
It is featured by collision resistance, corrosion resistance, high.temperature resistance, shape
retention and other performances
Measuring range: 0-100ug/L 0-20mg/L
Electrode material: 316L stainless steel
Temperature compensation resistor: 2.252K 22K Ptl00 Ptl000 etc
Sensor life: >3 years
Cable length: 5m (double shielded)
Detection lower Limit: 0.1ug/L(ppb)(20℃)
Measurement upper limit: 20mg/l(ppm)
Response time: ≤3min(90%, 20℃)
Polarization time: >8h
Minimum flow rate: 5cm/s; 515 L/h
Drift: <3%/month
Measurement error: <±1 ppb
Air current: 50-80nA   Note: Maximum current 20-25 uA
Polarization voltage: 0.7V
Zero Oxygen: <5ppb(60min)
Calibration intervals: >60 days
Measured water temperature:0~60℃
Measuring range: 0-100ug/L 0-20mg/L
Electrode material: 316L stainless steel
Temperature compensation resistor: 2.252K 22K Ptl00 Ptl000 etc
Sensor life: >3 years
Cable length: 5m (double shielded)
Detection lower Limit: 0.1ug/L(ppb)(20℃)
Measurement upper limit: 20mg/l(ppm)
Response time: ≤3min(90%, 20℃)
Polarization time: >8h
Minimum flow rate: 5cm/s; 515 L/h
Drift: <3%/month
Measurement error: <±1 ppb
Air current: 50-80nA   Note: Maximum current 20-25 uA
Polarization voltage: 0.7V
Zero Oxygen: <5ppb(60min)
Calibration intervals: >60 days
Measured water temperature:0~60℃

 

Applied to thermal power plants, power plant desalted water,.

boiler feed water etc places of trace oxygen content.


Product detail pictures:

China Wholesale Online Dissolved Oxygen Meter Manufacturers Pricelist - DOG-208F Industrial Dissolved Oxygen Sensor  – BOQU detail pictures


Related Product Guide:

Our aim is usually to give superior quality items at aggressive rates, and top-notch company to clients around the earth. We have been ISO9001, CE, and GS certified and strictly adhere to their good quality specifications for China Wholesale Online Dissolved Oxygen Meter Manufacturers Pricelist - DOG-208F Industrial Dissolved Oxygen Sensor – BOQU , The product will supply to all over the world, such as: Benin, Boston, Sri Lanka, We've a skilled sales team, they have mastered the best technology and manufacturing processes, have years of experience in foreign trade sales, with customers able to communicate seamlessly and accurately understand the real needs of customers, providing customers with personalized service and unique merchandise.
  • The product classification is very detailed that can be very accurate to meet our demand, a professional wholesaler.
    5 Stars By Barbara from Paraguay - 2018.09.19 18:37
    The customer service staff's attitude is very sincere and the reply is timely and very detailed, this is very helpful for our deal,thank you.
    5 Stars By Philipppa from Korea - 2018.06.12 16:22
    Write your message here and send it to us