1. Low detection limit, very suitable for power plant water feed, saturated steam and superheated steam silicon content detection and control;
2. Long life light source, using cold monochrome light source;
3. Historical curve recording function, can store 30 days of data;
4. Automatic calibration function, period arbitrarily set;
5. Support multi-channel measurements in water samples, optional 1-6 channels;
6. Achieve a maintenance-free, except adding reagents, guide standards.
| 1 | Measuring range: 0~20ug/L, 0~100ug/L, 0-2000ug/L, 0~5000ug/L(special) | 
| 2 | Accuracy: ± 1% F.S | 
| 3 | Repeatability: ± 1% F.S | 
| 4 | Stability: Drift ≤ ± 1% F.S / 24hours | 
| 5 | Response time: the initial response is 12 minutes | 
| 6 | Sampling period: about 10 minutes / Channel | 
| 7 | Watery conditions: Flow: > 100 ml / min Temperature: 10 ~ 45 ℃ Pressure: 10 kPa ~ 100 kPa | 
| 8 | Environmental conditions: Temperature: 5 ~ 45 ℃, Humidity: <85% RH | 
| 9 | Reagent consumption: Three kinds reagents, about 3 liters / month for each kind. | 
| 10 | Current output: 4 ~ 20mA arbitrarily set within this range, multi-channel meter, channel independent output | 
| 11 | Alarm output: normally open relay contacts 220V/1A | 
| 12 | Power supply: AC220V ± 10% 50HZ | 
| 13 | Power consumption: ≈ 50W | 
| 14 | Dimensions: 720mm (height) × 460mm (width) × 300mm (depth) | 
| 15 | Hole size: 665mm × 405mm | 