The incredibly rich projects administration experiences and a person to 1 service model make the substantial importance of organization communication and our easy understanding of your expectations for Water Conductivity Meter , Cod Meter , Dissolved Oxygen Transmitter , We feel that our warm and professional support will bring you pleasant surprises as perfectly as fortune.
China Wholesale Dissolved Oxygen Probe In Bioreactor Suppliers Factories - DOG-208F Industrial Dissolved Oxygen Sensor – BOQU Detail:

DOG-208F Dissolved Oxygen Electrode applicable for Polarography Principle

With platinum (Pt) as cathode and Ag / AgCl as anode
The electrolyte is 0.1 M potassium chloride (KCI).
The silicone rubber permeable membrane imported from U.S. serves as the permeable
membrane.
It has silicone rubber and steel gauze.
It is featured by collision resistance, corrosion resistance, high.temperature resistance, shape
retention and other performances
Measuring range: 0-100ug/L 0-20mg/L
Electrode material: 316L stainless steel
Temperature compensation resistor: 2.252K 22K Ptl00 Ptl000 etc
Sensor life: >3 years
Cable length: 5m (double shielded)
Detection lower Limit: 0.1ug/L(ppb)(20℃)
Measurement upper limit: 20mg/l(ppm)
Response time: ≤3min(90%, 20℃)
Polarization time: >8h
Minimum flow rate: 5cm/s; 515 L/h
Drift: <3%/month
Measurement error: <±1 ppb
Air current: 50-80nA   Note: Maximum current 20-25 uA
Polarization voltage: 0.7V
Zero Oxygen: <5ppb(60min)
Calibration intervals: >60 days
Measured water temperature:0~60℃
Measuring range: 0-100ug/L 0-20mg/L
Electrode material: 316L stainless steel
Temperature compensation resistor: 2.252K 22K Ptl00 Ptl000 etc
Sensor life: >3 years
Cable length: 5m (double shielded)
Detection lower Limit: 0.1ug/L(ppb)(20℃)
Measurement upper limit: 20mg/l(ppm)
Response time: ≤3min(90%, 20℃)
Polarization time: >8h
Minimum flow rate: 5cm/s; 515 L/h
Drift: <3%/month
Measurement error: <±1 ppb
Air current: 50-80nA   Note: Maximum current 20-25 uA
Polarization voltage: 0.7V
Zero Oxygen: <5ppb(60min)
Calibration intervals: >60 days
Measured water temperature:0~60℃

 

Applied to thermal power plants, power plant desalted water,.

boiler feed water etc places of trace oxygen content.


Product detail pictures:

China Wholesale Dissolved Oxygen Probe In Bioreactor Suppliers Factories - DOG-208F Industrial Dissolved Oxygen Sensor  – BOQU detail pictures


Related Product Guide:

We believe that long term partnership is a result of high quality, value added service, rich experience and personal contact for China Wholesale Dissolved Oxygen Probe In Bioreactor Suppliers Factories - DOG-208F Industrial Dissolved Oxygen Sensor – BOQU , The product will supply to all over the world, such as: South Africa, Slovakia, Mecca, We've got constantly insisted on the evolution of solutions, spent good funds and human resource in technological upgrading, and facilitate production improvement, meeting the wants of prospects from all countries and regions.
  • Reasonable price, good attitude of consultation, finally we achieve a win-win situation,a happy cooperation!
    5 Stars By Ella from New Zealand - 2018.06.30 17:29
    This is a reputable company, they have a high level of business management, good quality product and service, every cooperation is assured and delighted!
    5 Stars By Marcia from Denmark - 2018.02.21 12:14
    Write your message here and send it to us